Basics of X-ray and Mammographic Systems

Sponsored by
RSNA / AAPM

home


Image Geometry

Radiographic image unsharpness limited by
  • Focal spot size
  • Source-image distance (SID)
  • Object-image distance (OID)

Penumbra (partial shadow)

  • Less penumbra = less unsharpness

Minimizing unsharpness

  • Minimize focal spot size
    • Limits beam intensity
    • Lengthens exposure time
      • More potential for motion unsharpness
  • Maximize SID
    • Limits beam intensity
    • Lengthens exposure time
      • More potential for motion unsharpness
  • Minimize OID
    • Limited by
      • Thickness of table top
      • Space for grid & phototimer
      • Distance of body part of interest from where beam exits patient
home